AC/DC Power Tester

AC/DC Power Tester

4-新闻
News
April 21, 2025

AC/DC power tester

Model AIP8721P


AC/DC Power Tester


Product Features

√ AC/DC universal wide range design, DC ~ 1kHz bandwidth, better load adaptability.

√ Reliable, stable, small size, and light weight.

√ Fast measurement speed, with the fastest refresh rate of 0.25s.

√ Voltage, current ratio setting function, and easy to extend the measurement range.

√ Voltage, current, and power over the limit of sound and light alarm function, to meet the site quickly identify abnormal product requirements.

√ Standard RS-232 or optional RS-485 serial communication port, supports MODBUS communication, and is easy to achieve test automation.

√ An optional switching I/O function for remote control input and alarm signal output.


Typical application

√ Severe industrial field type tests.

√ High-temperature aging room test.

√ Standby power measurement of household appliances and commercial appliance production lines.

√ Power tool testing.

√ Lighting fixture testing.


Panel Schematic

Panel Schematic                      Panel Schematic

Boundary dimension



Specification


model number

AIP8721P

measurement channel

single phase (elec.)

measurement parameter

Voltage U, current I, active power P, reactive power Q, apparent power S, power factor λ, phase angle Φ, voltage frequency f, electrical energy Wh, electrical energy timing Time

Wiring Use

Voltage/current terminals are equipped with non-removable common ground shorting tabs, and internal connection method only.

Input Impedance

Voltage: approx. 2MΩ

Current: approx. 4mΩ (20A range), approx. 80mΩ (1A range)

Full-scale peak factor

1.6

Rated voltage range

600V

Voltage accuracy range

1V~600V

Rated range of current

1A/20A (AC/DC)

Current Accuracy Range

1mA~1A/20A

Voltage/current accuracy*

±(0.1%×displayed value+0.1%×range

Active power measurement range

2.2mW (PF=0.01) ~ 4.4kW (PF=1) @220V

Active power accuracy*

±(0.1% x display value + 0.1% x range)

Power Factor 

±(0.01 to 1)

Accuracy Rang

Frequency measurement range

DC, 45Hz≤f≤1kHz

Frequency measurement accuracy

±(0.1% x display value) (voltage amplitude should be greater than 10% of range)

Active power resolution

0.01W

Power Factor Resolution

0.001

Electrical energy range

0-99999MWh

Electrical energy accuracy

±(0.2 per cent x display value)

Extended uncertainty

Voltage/current/power/frequency/electrical energy uncertainty ≤ 0.30 per cent

Electricity Timing

999 hours, 59 minutes, and 59 seconds

Power Timing Accuracy

± (0.05% x display value)

Voltage/current ratio

0.1 to 1000.0

Data update cycle

0.25, 0.5, 1 [s]

Alarm function

5 groups, voltage, current, power upper/lower limit, threshold setting

control interface

RS-232 (optional RS-485), switching function (optional)

Overall dimensions

213mm (W, front moulding) x 88mm (H, front moulding) x 386mm (D, including terminal block)

Opening Size

210(W)×85(H) mm

Foot height

15 mm

Weight

Approx. 3kg

Power Consumption

Approx. 10VA

*[Precision conditions]

Pre-heating: more than 15 minutes after switching on

Temperature: 23±5℃, Humidity: 30%~75%RH, Power supply: AC 100V~240V, 45~65Hz

Temperature coefficient: 5~18℃ or 28~40℃, the error increases ± (display value × 0.06%)/℃